0102030405
HSIN-40Y Digital Sphygmomanometer Standard Calibrator
Description
HSIN-40Y Digital sphygmomanometer standard calibrator realizes intelligent detection by using microprocessor chip and special pneumatic components, It improves the detection efficiency and accuracy of sphygmomanometers (meters), reduces the labor intensity of personnel, and is suitable for detecting all kinds of sphygmomanometers (meters).The built-in automatic pressure generator can set the functional parameters of the instrument freely. After the indication error verification point is set, the instrument will automatically generate pressure to each verification point. After the pressure is adjusted and stabilized, a prompt tone will be sent out. At this time, the verifier will read out the indication value of the tested sphygmomanometer and compare it with the standard value to obtain the error of this point.
Feature
1.It has large screen LCD Chinese display and built in tips and operating instructions; It is simple and convenient to use.
2.It can inflate and adjust the air pressure to the required pressure point automatically and quickly with sound prompt.
3.It can verify the gas density and release rate items ,also process and calculate the display results automatically according to the metrological verification regulations.
4.It use imported pressure sensors and pneumatic components, its zero point can correct accuretely and automatically
5.It is built in overpressure protection safely and reliably.
6.The verification time is fast. It only takes 5 seconds to detect a pressure point.
7.The detection point and detection items can be selected quickly and stored once for ever
8.The special plastic spray mold opening case is beautiful and generous, safe and durable
Measuing Range: | 0~40KPa | Work Environment: | 20℃±10℃,50%-70% RH |
resolution Ratio: | 0.01Kpa | Power Supply: | AC220V/50Hz |
Verification Speed: | 3-5Seconds/Pressure Point | Instrument Weight: | 3Kg |
Accuracy Class: | 0.2% | Instrument Size: | (300*230*100)mm |
Verification Method: | Increase and decrease pressure automatically,Manual recording of indication error,Automatic verification of air tightness and outgassing rate items |
Verification Regulations
Comply with national metrological verification regulations
1.《JJG_692-2010_Noninvasive automatic sphygmomanometer》
2.《JJG 270-2008_Verification regulation of sphygmomanometers 》